- Perform a daily operational check for mechanical air removal in your steam sterilizers with this disposable test pack
- The ready-to-use test is easy to interpret
- Place the test pack on the lowest shelf of the sterilizer and expose the pack for 3.5 minutes at 132 degrees C
- Simply remove the test sheet from the center of the test pack following test cycle
- The pattern on the test sheet turns completely from blue to black to indicate a satisfactory cycle, the test´s ink chemistry retains its vivid blue color if an air leak is present
Especificaciones
Especificaciones
-
ApplicationSterilization Bowie-Dick Test Pack
-
Indicator ChangeBlue to Black
-
Sterilization CompatibilitySteam
-
Temperature Range270°F (132°C)
-
Time to Results3-1/2 Minute Exposure Time
-
TypeType 2